|   
           Thermal Dilatometric   
          Analysis (TDA), often called "dilatometry", measures   
          the dimensional change of a material (ceramics, glasses, metals,   
          composites, carbon/graphite, minerals, polymers, and others) as a   
          function of temperature. This test determines both reversible and   
          irreversible changes in length (expansion and shrinkage) during   
          heating and cooling, and pinpoints where reactions occur that cause   
          expansion or contraction. Samples are quickly and easily measured for   
          determining firing ranges and firing schedules, measuring thermal   
          expansion ranges for glaze fits, and measuring thermal expansion   
          ranges for R&D, QC or product certification. Orton dilatometers   
          are used for ASTM E-228 testing.   
          A. Characteristics or   
          Properties Measured   
          Coefficient of Thermal Expansion (CTE),      
          softening point, glass transition temperature, curie point,      
          crystalline transformation, phase transition, shrinkage, warping,      
          bloating, sintering rate, isothermal creep, stress relaxation.      
                
          The test results are a graph of the TDA signal (converted to percent      
          length change) on the Y-axis plotted versus the sample temperature in      
          ¡ÆC on the X-axis. Sample graphs of enhanced output are shown below.      
               
                 
                   
                | Examples of      
                  Applications |      
                    
                   
                       
                  Ceramics - ASTM E-228 |      
                       
                  Metals MIL I-23011C Class 7 |      
                    
                 
                
                
                
               
                 
                   
                | Range of TDA Test Conditions: |      
                     
                       
                    - -150¡ÆC to +1,600¡ÆC     
                    
 - Ambient, Inert, Reducing Atmospheres     
                    
 - Simple Heat-up and Cool-down, Iso-thermal Holds,      
                      Programmed Thermal Cycles
      
                        
                 |      
                    
                 
                
                 
          B. Standard Dilatometers      
               
                 
               
                 
                   
                | Standard,      
                  Single Sample, Horizontal Dilatometers |      
                    
                  |      
                DIL      
                  2010 B |      
                DIL      
                  2010 C |      
                DIL      
                  2010 STD |      
                DIL      
                  2012 STD |      
                DIL      
                  2016 STD |      
                    
                   
                | Temperature Range |      
                RT to 1,000¡ÆC |      
                RT to 1,000¡ÆC or      
                  -170¡ÆC to +300¡ÆC |      
                RT to 1,000¡ÆC |      
                RT to 1,200¡ÆC |      
                RT to 1,600¡ÆC |      
                    
                   
                | Furnace |      
                Kanthal - Tube |      
                Nichrome - Split Shell      
                  Cryogenic Chamber |      
                Kanthal - Tube |      
                Kanthal - Tube |      
                Silicon Carbide - Tube |      
                    
                   
                | Thermocouple |      
                Type "N" |      
                Type "N" |      
                Type "S" |      
                Type "S" |      
                Type "S" |      
                    
                   
                | Sample Holder and Probe Rod |      
                Fused Quartz |      
                Fused Quartz |      
                Fused Quartz |      
                High Alumina |      
                High Alumina |      
                    
                   
                | Sample Size (max) |      
                50 mm long by      
                  20 mm diameter |      
                100 mm long by      
                  10 mm diameter |      
                100 mm long by      
                  20 mm diameter |      
                50 mm long by      
                  20 mm diameter |      
                50 mm long by      
                  20 mm diameter |      
                    
                   
                | Contact Load |      
                113 grams |      
                Adjustable      
                  4 grams min. |      
                Adjustable      
                  4 grams min. |      
                Adjustable      
                  4 grams min. |      
                Adjustable      
                  4 grams min. |      
                    
                   
                | Temperature      
                  Control |      
                Orton      
                  Multi-segment Controller |      
                    
                   
                | Data Acquisition |      
                Orton      
                  On-board Computer |      
                    
                   
                | Data Analysis |      
                Orton      
                  Analysis Software (Windows 95/98/2000 Based) |      
                    
                   
                | Computer      
                  Interface |      
                RS232      
                  Cable |      
                    
                   
                | Controlled      
                  Atmosphere Option |      
                Not      
                  Available |      
                Yes |      
                Yes |      
                Yes |      
                Yes |      
                    
                   
                | Power      
                  Requirements |      
                120      
                  VAC, 15 amp, 50/60 Hz |      
                120      
                  VAC, 15 amp, 50/60 Hz |      
                120      
                  VAC, 15 amp, 50/60 Hz |      
                120      
                  VAC, 15 amp, 50/60 Hz |      
                240      
                  VAC, 205 amp, 50/60 Hz |      
                    
                 
                
          *Descriptions and specifications are subject to      
          change without notice.     
          ¡¡      
                 
               
                 
                   
                  |      
                     
                   Other Orton Dilatometers      
                  • Vertical      
                  • 2 Sample      
                  • Multiple Sample      
                  • Non-contact Laser      
                  • Rapid turnaround      
                  • Quench (metallurgical)      
                 |      
                  |      
                    
                 
                
          Contact Orton for more      
          details.      
          C. Computer Analysis      
          Every Orton dilatometer is supplied      
          with the software to add to the user's PC in order to acquire, save      
          and analyze the data generated by the dilatometer. The Orton      
          Dilatometer Software is a Visual Basic executable routine written for      
          Windows 95/98/2000 based personal computers. It can be used to monitor      
          the dilatometer test in real time, or can be used to examine the test      
          data after the run. The software imports the data through the RS232      
          interface, and stores it on the hard drive for immediate or      
          post-testing analysis. The software enables the user to:      
                
          View the dilatometer data in a variety of      
          presentations      
               
            - percent linear change (PLC) vs.      
              temperature     
            
 - percent linear change (PLC) vs. time     
            
 - first derivative of the percent linear      
              change (DCE) vs. temperature     
            
 - first derivative of the percent linear      
              change (DCE) vs. time     
            
 - percent linear change (PLC) and first      
              derivative of the percent linear change (DCE) vs. temperature     
            
 - percent linear change (PLC) and first      
              derivative of the percent linear change (DCE) vs. time
      
                
          Perform a variety of analyses      
               
            - calculate the coefficient of thermal      
              expansion (CTE) between specified temperatures, or a series of      
              specified temperatures     
            
 - calculate the average coefficient of      
              thermal expansion from room temperature to a specified temperature      
              at a specified temperature increment     
            
 - determine transition temperature     
            
 - determine softening temperature     
            
 - locate alpha-beta quartz transition      
              Export the data in a text file format for independent analysis or      
              archiving purposes
      
                
               
          D. Additional      
          Information on TDA      
          All materials expand and contract as a function      
          of temperature. For two materials to adhere to each other, such as      
          glass to metal seals, metalizations to substrates, and glazes to      
          bodies, their respective thermal expansion characteristics must be      
          known, matched, and controlled. Thermal Dilatometric Analysis (TDA),      
          often called Dilatometry, measures the amount of dimensional change of      
          a material (ceramics, glasses, metals, composites, carbon/graphite,      
          minerals, plastics, and others) during a controlled thermal cycle.      
          Dilatometry measures the normal expansion and contraction of a      
          material, including its reversible phase changes. This procedure also      
          measures the irreversible changes in length that are the result of      
          decompositions, phase transformations, and other chemical reactions,      
          and helps identify the temperature ranges of those events and      
          reactions. Such testing is helpful when trying to control the thermal      
          expansion characteristics of various lots of materials, and in      
          determining drying and firing schedules.      
               
                 
                   
                | Principle of Operation |      
                  |      
                    
                 
                
          The sketch above shows the concepts of a      
          dilatometer. A sample specimen is placed between the end of the sample      
          holder and the end of the movable probe rod, and the furnace is heated      
          according to a pre-programmed thermal cycle. As the sample temperature      
          changes (as recorded by the sample thermocouple), the sample expands      
          (pushing against the probe rod) or shrinks (pulling away from the      
          probe rod). The probe rod transmits the amount of sample movement to      
          an electronic displacement sensor located outside of the heated      
          chamber. The displacement sensor generates an electronic signal      
          corresponding to the positive or negative change in sample length and      
          continuously sends the signal to the computer. The computer converts      
          the signal to the percent of length change (%DL) and saves it along      
          with the elapsed time and the sample temperature. The basic TDA curve      
          is generated by plotting the percent of length change (%DL) on the      
          Y-axis against the sample temperature.      
               
                 
                   
                | Horizontal      
                  Dilatometer |      
                  |      
                  |      
                    
                 
                
          The photos above are a horizontal      
          dilatometer with the furnace moved away to expose the sample holder,      
          and a close-up view of the sample holder. The photo on the right shows      
          how the sample is positioned between the end of the sample holder and      
          the probe rod. After positioning the sample in the sample holder, the      
          furnace is moved horizontally to surround the sample and sample      
          holder.      
          The probe rod extends from the end of the      
          sample, throught the sample holder tube, and connects to the      
          displacement sensor assembly outside the furnace. The probe rod is      
          spring loaded outside the furnace to keep it in constant contact with      
          the sample, even when shrinking.      
          The main advantage of the horizontal system      
          is the uniform temperature zone for the sample. Most dilatometer tests      
          are performed with a horizontal unit.      
          Vertical Dilatometer      
                
          For larger samples, such as structural clay      
          bodies, a vertical  dilatometer      
          is used. The sample is placed into the furnace and the vertical probe      
          rod is lowered to contact the sample (as shown in the photo at the      
          right). The furnace is heated according to the pre-programmed thermal      
          cycle. As the sample temperature changes, the sample expands, pushing      
          up on the probe rod, or shrinks, pulling away from the probe rod. The      
          probe rod is vertically suspended and counterweighted so that gravity      
          keeps it in constant contact with the sample. The probe rod transmits      
          the amount of sample movement to the electronic displacement sensor      
          located overhead and outside the furnace.      
                 
          E. Frequently Asked Questions:      
          Percent Length Change (PLC)      
          and Coefficient of Thermal Expansion (CTE): Percent Linear      
          Change (PLC) is the amount of expansion or shrinkage expresses in      
          percentage of an initial length. A standard TDA curve is usually the      
          PLC on the Y-axis and the temperature on the X-axis. The thick, black      
          line in the graph below is a typical TDA curve of glass.      
          Temperature: The temperature range for a      
          dilatometer is determined by the type of heating element or heating      
          system used. The standard Orton dilatometers are made for one of the      
          following temperature ranges:      
               
            - -170¡ÆC to +300¡ÆC (Cryogenic cooling      
              chamber and Ni-chrome Heating Element with Fused Quartz Sample      
              Holder and Probe Rod)     
            
 - Room Temperature to 1,000¡ÆC (Kanthal      
              Heating Element with Fused Quartz Sample Holder and Probe Rod)     
            
 - Room Temperature to 1,200¡ÆC (Kanthal      
              Heating Element with High Alumina Sample Holder and Probe Rod)     
            
 - Room Temperature to 1,600¡ÆC (Silicon      
              Carbide or Platinum Heating Element with High Alumina Sample      
              Holder and Probe Rod)     
            
 - Room Temperature to 1,700¡ÆC      
              (Molybdenum Disilicide Heating Element with High Alumina Sample      
              Holder and Probe Rod)
      
                
          Thermal Cycle: The      
          most commonly used thermal cycle for dilatometry is a simple,      
          controlled heat-up rate of 3¡ÆC per minute from ambient temperature to      
          the maximum temperature, then the test is terminated. For testing      
          materials that experience irreversible reactions or when quartz      
          transitions are critical, the thermal cycle can be extended to include      
          a cool down segment. For developing firing schedules or examining what      
          happens during a certain firing schedule, the dilatometer can be      
          programmed to follow an actual production schedule that contains      
          multiple ramps and soaks. The Orton dilatometer can be programmed for      
          simple cycles, or up to a 20 segment thermal cycle.      
          Heat-up Rate: A      
          sample does not absorb heat instantaneously, so it does not expand or      
          shrink instantaneously. Since a finite amount of time is required for      
          a sample to come to an equilibrium temperature and expansion/shrinkage      
          condition, some thermal expansion measurements are made by holding the      
          furnace temperature constant until the sample reaches an equilibrium      
          temperature and length. These static condition (isothermal) tests take      
          a lot of time, and are generally performed at only a few temperatures.      
          ¡¤ To save time and generate more information over a broader range of      
          temperatures, most thermal expansion measurements are taken while the      
          sample is being heated during a dynamic heat up. The thermal      
          conductivity, size, and geometry of the sample will influence how      
          quickly the sample can absorb heat and change length. With fast heat      
          up rates, there is a tendency for the sample temperature to lag behind      
          the furnace temperature, and the corresponding change in length to lag      
          as well. With slow heat up rates, the amount of thermal lag and length      
          change lag are much less, but the tests can be very long. Over the      
          years the common industrial practice has evolved to a compromise rate      
          of 3¡ÆC per minute. This keeps the amount of lag to a minimum and the      
          testing time practical. ¡¤ The important factor is to be consistent in      
          the heat up rate used, and to be consistent with the sample size and      
          geometry.      
          Sample Size:      
          The maximum sample size for the horizontal dilatometer is 2" long      
          by 3/8" diameter or square. Samples longer than 2" will      
          extend beyond the isothermal zone, and larger diameter samples will      
          not fit into the sample holder. The ends of samples must be flat,      
          parallel, and perpendicular to the length axis. ¡¤ The ultimate sample      
          length depends upon the total amount of expansion or the total amount      
          of shrinkage expected. The displacement sensor on the standard Orton      
          dilatometer has a total linear movement of 5.0 millimeters (0.200      
          inches). Orton prefers to use only the center half of this range, so      
          total sample movement (expansion or shrinkage) is 2.5 millimeters      
          (0.100 inches). Consequently, low expansion/shrinkage samples should      
          be long, and high expansion/shrinkage samples can be shorter. The      
          following table shows the maximum percent expansion / shrinkage of      
          different length samples.      
               
                 
                   
                |      
                        
                    Sample Length      
                    millimeters     
                        
                 |      
                     
                        
                    Sample Length      
                    inches     
                        
                 |      
                     
                        
                    Percent      
                    Expansion/Shrinkage     
                        
                 |      
                    
                   
                |      
                        
                    5      
                    10      
                    15      
                    20      
                    25      
                    30      
                    35      
                    40      
                    45      
                    50     
                        
                 |      
                     
                        
                    0.197      
                    0.394      
                    0.591      
                    0.787      
                    0.984      
                    1.181      
                    1.378      
                    1.575      
                    1.772      
                    1.969     
                        
                 |      
                     
                        
                    50.00%      
                    25.00%      
                    16.67%      
                    12.50%      
                    10.00%      
                    8.33%      
                    7.14%      
                    6.25%      
                    5.56%      
                    5.00%     
                        
                 |      
                    
                 
                
                
          Sensitivity: (To be added later)      
          Atmospheres: Tests      
          are normally performed in ambient air. By adding the controlled      
          atmosphere option to the Orton dilatometer, inert and reducing      
          atmospheres, as well as vacuums are possible. There are several      
          concerns when running tests in controlled atmospheres:      
               
            - The atmosphere tube surrounding the      
              sample is a thermal barrier, and creates a larger thermal lag      
              between the heating elements and the sample. Faster heating rates      
              create an even larger thermal lag. Slower heating rates may be      
              desirable.     
            
 - Flowing a gas over the sample will cool      
              the sample, and create an even greater thermal lag between the      
              sample and the heating elements. Flow rates should be a low as      
              possible to minimize this affect.     
            
 - Reducing atmospheres, such as hydrogen,      
              will significantly degrade platinum thermocouples at elevated      
              temperatures. Small amounts of moisture with the hydrogen will      
              increase the rate of platinum degradation. Using "dry"      
              hydrogen will reduce the moisture attack. Using a surrogate gas,      
              such as dry helium, may be reducing enough, and will retard      
              platinum deterioration. A last alternative is to use a shielded      
              thermocouple to totally avoid platinum degradation. However, a      
              shielded thermocouple introduces another thermal lag into the      
              test.     
            
 - A vacuum is an excellent thermal      
              barrier to heat transfer. There is no gas for conduction and      
              convection. Radiation does not become an effective heat transfer      
              mechanism until elevated temperatures.
      
                
          ¡¡  |